The AC1036 is designed to facilitate the calibration of any CAP-TT01, MTT01, CAP-TA (CTA), CAP-T (CT) Cap Torque Tester. The kit contains a complete set of attachments required to mount the tester to a bench, as well as the necessary brackets and cable to mount weights, as follows:
- Right-angle bench mounting bracket
    
- Cable with looped ends
- Attachment, cable to tester
- Set of hardware
Weights are not included. Suggested source for calibration weights:
    Rice Lake Weighing Systems - www.ricelake.com
    Troemner Precision Weights - www.troemner.com
CAP-TT01 torque testers must be calibrated at full scale in English units only. Refer to the CAP-TT01 user's guide for complete calibration instructions. The following weight combinations are recommended:
CAP-TT01-12: 
     3.6 lb, as follows:
    1 lb hanger weight- qty. 1
    2 lb slotted weight - qty. 1
    0.5 lb weight - qty. 1
    0.1 lb weight - qty. 1
CAP-TT01-25: 
     7.5 lb, as follows:
    1 lb hanger weight- qty. 1
    5 lb slotted weight - qty. 1
    1 lb weight - qty. 1
    0.5 lb weight - qty. 1
CAP-TT01-50: 
  15 lb, as follows:
    1 lb hanger weight- qty. 1
    10 lb slotted weight - qty. 1
    2 lb slotted weight - qty. 2
CAP-TT01-100: 
 
    30 lb, as follows:
    1 lb hanger weight- qty. 1
    20 lb slotted weight - qty. 1
    5 lb slotted weight - qty. 1
    2 lb slotted weight - qty. 2

 
				

 
 
					 
					